Ø Chinese Name:气相色谱仪
Ø Model: Agilent 6820
Ø Instrument number: 20047182
Ø Manufacturer: America Agilent
Ø Acquisition Price:¥140000
Ø Acquisition Date:2004-06-19
Ø Locate: Room101, 24thbuilding,
Wing A Tianjin University
Ø Specification:
u Column oven:
Column Temperature: Room temperature: +4°C~425°C;
Temperature set the resolution:1°C,maximum heating rate :75°C/min;
Maximum run time: 999.99min, temperature programme: 5level;
Temperature precision: Room temperature fluctuations per 1℃, column oven temperature fluctuations < 0.01℃; cooling rate:350℃~50℃/6.5min.
u Heating range: Except the column temperature oven , there are six independent heating area (2 injection ports, and 2 detectors and two auxiliary areas) assisted the maximum temperature: 400°C;
u Split/splitless injection port: Max temperature: 400°C; Pressure range: 0 ~ 30psi;
u FID detector: Can be used in capillary column and packed column, high temperature: 425°C;Minimum detection limit: <5 pg carbon/sec.
Ø Main Application:
u Environmental protection: toxic trace analysis, detection and study of atmosphere water pollution; Biochemistry: clinical application, pathology and toxicology research;Food fermentation: analysis of trace components in microbiological drinks;
u Chinese and western medicine: analysis of intermediates and products;
u Oil processing: analysis and control the compositions in petrochemical, petroleum geology, oil control and analysis, and study on the empty ore, etc;
u Organic chemistry: research and production control in organic synthesis in the field of components; Sanitary inspection: the analysis and research of the labor protection public hazard detection;
u Advanced science: military control testing and research.